$482.00

  • Measurement resolution 0.1 lx (0.01 fc).
  • High accuracy and fast response
  • Data-hold function for holding measuring values
  • No need for correction factors calculating for different light sources thanks to a very good fit spectral sensitivity guaranteeing the correct measurement of intensity of the illumination regardless of the nature of radiation
  • Peak-hold function for tracing the peak signal of light pulse with duration longer than 0.4 s for LXP-2) and less than 1 s
  • Capable of selecting measuring mode in Lux or FC scale alternatively
  • Auto power off (5,10 or 15minutes) or disable AUTO power off
  • Maximum and minimum measurements
  • Relative reading
  • Easy to read large backlit display

Available on backorder

Category: Tag:
The most important features:

  • Measurement resolution 0.1 lx (0.01 fc).
  • High accuracy and fast response.
  • Data-hold function for holding measuring values.
  • No need for correction factors calculating  for different light sources thanks to a very good fit spectral sensitivity guaranteeing the correct measurement of intensity of the illumination regardless of the nature of radiation,
  • Peak-hold function for tracing the peak signal of light pulse with duration longer than 0.4 s for LXP-2) and less than 1 s,
  • Capable of selecting measuring mode in Lux or FC scale alternatively.
  • Auto power off (5,10 or 15minutes) or disable AUTO power off.
  • Maximum and minimum measurements.
  • Relative reading.
  • Easy to read large backlit display
  • USB output connect with PC.
  • Four measurement ranges.
  • Built-in memory – 99 result, with can be checked in the meter or PC.
  • 16000 values records data logger (traceable only in PC)

 

Illuminance measurement probe LP-1

Display
Range
[lx]
Resolution
[lx]
Spectral
uncertainty
Accuracy
0…399,9 0,1 f1‘ < 6% ±(5 % + 5 digits)
400…3999 1
4,00 k…19,99 k 0,01 k

 

Display
Range
[fc]
Resolution
[fc]
Spectral
uncertainty
Accuracy
0…39,9 0,01 f1‘ < 6% ±(5 % + 5 digits)
40,0…399,9 0,1
400…1999 1

– display result in lx or fc
– meter class B

Rated operational conditions:

  • operating temperature and humidity: 0…50°C; 0% to 80% RH,
  • storage temperature and humidity: -20…70°C; 0% to 70% RH,

Other technical data:

  • display: 3¾ digit LCD with high speed 40 segment bargraph,
  • over range indication: „0L” displayed,
  • spectral response: CIE photopic (CIE human eye response curve),
  • cosine response (f2’) : ±3%,
  • sampling rate: 1,3 times/sec,
  • power supply: battery 9V or rechargeable battery 8,4 V,
  • photo detector: one silicon photo diode and spectral response filter,
  • memory: 99 results,
  • photo detector lead length: approx. 100cm,
  • photo detector dimensions: 115 × 60 × 20mm,
  • meter dimensions: 170 × 80 × 40mm,
  • weight: 390g,
  • data transmission: USB cable.